Abstract

The Günther Tulip inferior vena cava (IVC) filter is designed for transjugular retrieval with use of a sheath and snare device. This report describes a technique for removal of difficult-to-retrieve Günther Tulip IVC filters when the standard method fails. In a series of four patients, each with an IVC filter refractory to capture by snare alone, the use of a snare-over-guide wire loop technique succeeded in retrieving the filter in all cases.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.