Abstract

The authors report a case of inferior vena cava filter arm perforation, followed by fracture and migration to the right ventricle causing chest pain and nonsustained ventricular tachycardia. Review of abdominal computed tomographic (CT) scans show the two filter arms that fractured and migrated had perforated the vena cava 2 years previously. Microscopic evaluation of the retrieved filter and limbs revealed bending metal fatigue at the fracture sites. This case and review of the literature suggest a causal relation between Bard Recovery filter arm perforation and subsequent fracture and migration. Percutaneous retrieval of filters with arm fracture or arm migration is recommended.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.