Abstract

The problem of retrieving the time dependence of the neutron energy spectrum from the detector current data in Faraday-cup experiments is studied. The time-dependent energy spectrum can be retrieved only when the neutron source in question meets a certain criterion. A detailed analytic formulation for the folding process is presented. The method that can be used to unfold the data in the retrievable case is explained in detail. 2 figures.