Abstract

The pixel by pixel calibration of a scientific CCD camera allows Poissonian statistics of the spatial fluctuations of an uniform enlightening to be retrieved in the full range of the camera dynamic. The procedure works efficiently for thermal as well as for laser sources, provided that the wavelength and the coherence properties of the source are chosen in order to avoid the formation of equal thickness fringes in the chip (etaloning effect). Calibration allows also the comparison at the shot noise level of images recorded at different places on the chip.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.