Abstract

Multi-focal intra-ocular or contact lenses, intended to increase depth of focus, conventionally have annular zones of additional refractive power, generating wavefront rings of coaxial spherical surfaces. It is, however, possible to influence depth of focus by changing not only the curvature of the wavefront, i.e., refractive power, in the annulus, but also the tilt, i.e., circularly symmetrical linear radial deviation imposed on the spherical wavefront. Employing the example of a single annulus bifocal, retinal image light distributions in the two regimes are calculated, using standard diffraction theory. Four measures of retinal image quality in through-focus scans show that plus power additions and wavefront tilts operate almost interchangeably. In testing these lenses, attention needs to be paid to the detailed operating characteristics of measuring devices of the Shack-Hartmann type to ensure that their grain and precision is compatible with the framework of the analysis.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.