Abstract
To investigate the maximum allowable deviation of retention time (RT) or relative retention time (RRT) between the common poisons (drugs) and standard solvent by gas chromatography-mass spectrometry (GC-MS). After pretreatment with liquid-liquid extraction, four common poisons (drugs)-dichlorvos, phorate, diazepam and estazolam-were detected by full scan mode GC-MS. RT and RRT were analyzed according to combined uncertainty and expanded uncertainty. The expanded uncertainty of RT and RRT were 6.0×10-4-14.1×10-3 and 2.5×10-6-5.9×10-5 (k=3), respectively. The RT of poisons (drugs) was relatively stable in blood samples with different mass concentrations. Among dichlorvos, phorate, diazepam and estazolam, the absolute deviation and relative deviation of RT were ≤0.03 min and ≤0.4%, respectively, and those of RRT were ≤0.003 min and ≤0.3%, respectively. The maximum allowable deviations of RT and RRT for common poisons (drugs) in blood samples are recommended to be ±0.05 min and ±0.5%.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.