Abstract
The influence of mobile phase composition on the retention of selected test analytes in different normal‐phase TLC systems has been studied. A novel adsorption model for an accurate prediction of the analyte retention in the thin‐layer chromatography with binary mobile phase has been proposed. The performance of the model was compared with the retention models reported in the literature. All the models were verified for different TLC systems by three criteria: the sum of squared differences between the experimental and theoretical data, the approximation of the standard deviation, and the Fisher test.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of Liquid Chromatography & Related Technologies
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.