Abstract

This study presents the results of the verification of the method according to working procedures of the Laboratory of External Dosimetry (LDE) located at the Laboratory of Radiation Physics and Metrology (LAF-RAM), which is the only Individual Monitoring service provider in Nicaragua. The test is considered as validation of the method since the LAF-RAM follows a Quality Management System. The chosen performance testing according to IEC 62387:2012 are Coefficient of variation, Non linearity, Radiation energy and angle of incidence, Dosimeter droop, Light exposure (reader), Over-response of radiation incidence and Build up, fading, self-irradiance and natural radiation response. The results in general show that all test are in compliance with the requirements except for some temporal conditions in Buildup, fading, self-irradiance test.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.