Abstract

Main textThis is the final report for the EURAMET.RI(II)-K2.Ho-166 key comparison of the activity of the radionuclide 166Ho per unit mass of aqueous solution which was organized in 2017 by Czech Metrology Institute (CMI) within the joint research project "Metrology for clinical implementation of dosimetry in molecular radiotherapy" (MRTDosimetry; 15HLT06) of the European Metrology Programme for Innovation and Research (EMPIR). The key comparison reference value of (2623.6 ± 3.6) kBq·g-1 was determined from the results of three participants at the reference time of 1 December 2017 00:00 UTC. The details of the comparison are presented and analysed in this paper.To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database https://www.bipm.org/kcdb/.The final report has been peer-reviewed and approved for publication by the CCRI, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.