Abstract

Long-term results of total hip arthroplasty (THA) using highly cross-linked polyethylene (HXLPE) and metal femoral head with more than 10 years of follow-up have already been reported. However, most studies included results with a head size of 28 mm that could affect wear rates. The aim of this study was to evaluate the results of 36-mm metallic femoral heads on first-generation HXLPE in patients less than 60 years of age with a minimum follow-up of 10 years. Retrospective analysis included 54 cases from 47 patients. The mean age at the time of surgery was 47.22 years and the mean follow-up period was 131.04 months. Porous-coated cementless acetabular cups (Trilolgy®; Zimmer Inc., Warsaw, Indiana, USA) and HXLPE acetabular liners (Longevity®; Zimmer Inc.) were used for all cases. Acetabular cup abduction angles, anteversion angles, and wear rates of liner were measured using polyWare pro 3D distal version 5.10. The average modified Harris hip score at the final follow-up was 88.48 (range 80-96). Average Merle d'Aubigne and Postel score was 15.57 (range 14-18). There was no acetabular cup or femoral stem failing due to aseptic loosening. The average steady-state wear rate determined using radiographs taken at 1 year postoperatively and at the latest follow-up was 0.053 ± 0.025 mm/year. There were no statistically significant differences in liner wear rate with respect to age, variety of the femoral stem, or liner thickness. Results of THA with 36-mm metallic femoral heads on first-generation HXLPE in patients less than 60 years of age were satisfactory.