Abstract
Outbreaks of infections transmitted by duodenoscopes have prompted changes to duodenoscope cleaning protocols. However, little is known about effectiveness of clinical cleaning practices. We present culture data after duodenoscope manufacturer-recommended high-level disinfection (HLD) and quarantine. Our institution adopted a combination of manufacturer-recommended cleaning with the Centers for Disease Control and Prevention-recommended culture and quarantine in 2015. Duodenoscopes (models TJF-Q180V, TJF-160, and PJF-160; Olympus, Center Valley, Penn, USA) underwent HLD according to the manufacturer's reprocessing manual protocols after use. Two culture specimens were then obtained using a sterile brush from the distal tip, including elevator mechanism, and by flushing sterile water through the working channel. Duodenoscopes were quarantined until cultures resulted. Positive cultures were defined as >10 colony-forming units (CFUs) of low-concern organisms or any CFUs of high-concern organisms according to Centers for Disease Control and Prevention recommendations. If either culture specimen was positive, the process was repeated until cultures were negative. Data were collected from December 2015 to July2018. There were 140 instances of duodenoscope cleaning with 280 specimens. Twenty-eight of 234 (12%) initial culture specimens (18% of cultured duodenoscopes) were positive. Five of 36 (14%) second cultures were positive. Two of 8 (25%) third cultures were positive. Ninety-eight percent of organisms cultured were gram positive. In 8 instances both culture methods (brushing and flush) were positive; otherwise, only 1 method was positive. In 11 instances (8%) duodenoscopes were removed from quarantine before final culture results. No patient had infections related to ERCP. Eighteen percent of duodenoscopes had a positive culture after initial HLD. Only 3% were gram-negative bacteria. Repeated HLD was 86% and 75% effective at eliminating initial and repeat positive cultures, respectively. Initial HLD per manufacturer recommendations is not always effective at eliminating bacterial contamination. Additional steps are necessary to decrease risks of duodenoscope-transmitted infections.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.