Abstract
A modified method of combined SAFT (C-SAFT) for restoring the images of reflectors is considered; it allows the multiple reflection of a pulse from the boundaries of the wall of a cylindrical inspection object. To verify the efficiency of the proposed algorithm, images of a crack were restored from echo signals that were calculated using the CIVA software package, which is intended for modeling the propagation and scattering of ultrasonic pulses. It was shown in a model experiment that the consideration of changes in the pulse phase during reflection from the inspection-object boundaries at various incident angles of an S-wave in the image-restoration algorithm increases the frontal resolution by more than two times. The consideration of five reflections from the inspection-object boundaries made it possible to obtain images of reflectors by the M-C-SAFT method using many acoustic schemes. The images allow one to determine the type of defects, as well as their dimensions and location in the depth of the wall of a pipeline that is 720 mm in diameter.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.