Abstract

Glow curves of LiF TLDs (TLD-100) were measured for 10-40 keV monoenergetic photons from synchrotron radiation, and the energy and dose responses were evaluated based on the glow curve data. The ratio of measured to calculated energy response was almost unity for the lower temperature region of the glow curve including the main peak, and was about 2.0 for the higher one. Dose responses to 10 keV 30 keV X rays, and 60 Co gamma rays showed different supralinearity. The maximum value of supralinearity decreased with decreasing photon energy, that is, with increasing the average linear energy transfer of incident photons

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.