Abstract

The responses of a commercial diamond detector (type 60003, PTW-Freiburg) to several heavy ions were examined. The responses to heavy-ion beams reached stable levels with relatively small pre-irradiation doses compared to photon-beam irradiations. The responses also reached stable levels with a smaller pre-irradiation dose when the dose rate of the He beams was increased. A total accumulated dose of about 5 Gy was required for the pre-irradiation dose of heavy-ion beams. No angular dependence of the detector responses was observed within a deviation of 5%. The dose-rate dependence of the detector responses to heavy-ion beams was far smaller than that to gamma rays. The decrease in the response was within 0.9%, with a variation from 0.88 to 18.2 Gy min−1 in the carbon beam. We examined the LET dependence of the diamond detector responses using various kinds of heavy-ion beams. The responses had particle dependence in addition to LET dependence. The responses decreased more with higher LET particles and decreased less with large-Z particles. We proposed a gradual-saturation model based on the track structure under several simple assumptions to explain the LET and particle dependences of the response.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call