Abstract

The Box-Behnken response surface design together with the individual desirability functions were used to develop the new and greenish sample preparation procedure of coffee brews prior to their multielement (Ba, Ca, Cu, Fe, K, Mg, Mn, Na, Sr and Zn) analysis by inductively coupled plasma optical emission spectrometry (ICP OES). The developed procedure required only 2-fold dilution of the samples with a 1.8molL-1 HNO3 solution and then, the sonication of the resulting samples solutions for 8min at room temperature. The proposed method was precise (0.6-7.5% as RSD), true (relative errors changing from -5.2% to +4.6%) and guaranteed the limits of detection (LODs) of the studied elements between 0.1 and 5ngg-1. Finally, this simplified ICP OES based method was applied for the multielement analysis of brews of different Arabica coffees as well as those prepared with seldomly reported devices, i.e., dripper, slow dripper, French press, aeropress and syphon.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.