Abstract

In this study, thermoluminescence detectors (TLD)-600 and TLD-700 were used under different conditions to study neutron dosimetry for its application in medical dosimetry as albedo dosimeters. 6 LiF has a high cross-section for neutrons and is more sensitive to neutrons than to gamma radiation. Conversely, TLD-700 showed a better response to gamma radiation. Therefore, to obtain the response for neutrons, the responses for TLD-600 and TLD-700 were subtracted. A cadmium sheet was used to absorb incident thermal neutrons so that detector measured only backscattered neutrons from the albedo dosimeter. A Perspex sheet was used as a moderator to thermalize the fast neutron source Am-Be used in the experiment. CR-39 was used to detect fast neutrons and act as an albedo dosimeter when covered with a 10 B disc. Both TLD and CR-39 dosimeters were successful in detecting scattered neutrons in radiotherapy rooms.