Abstract

The response of a TL dosemeter to beta radiation is heavily influenced by the absorption of the radiation in the dosemeter thickness. As a consequence, the assessment of skin dose-depends either on the execution of a calibration with a beta field of the same characteristics as that to be monitored or on the knowledge of depth-dose distribution in the dosemeter. Depth-dose distributions have been experimentally estimated for optically transparent dosemeters in a number of irradiation geometries and with sources of different configuration and energy. General algorithms based on the point-source function of Loevinger (1956) have been developed, by which the response of TL dosemeters can be evaluated and skin dose correction factors derived. TLD responses to beta radiation calculated by the present method are in sound agreement with other author's measurements.