Abstract

The measured space point resolution of a D0 Muon Chamber is /+-/0.31 mm perpendicular to the anode wire and 2.7 mm parallel to the wire. A voltage change of 1 kV, which changes the gas gain by a factor of 50, only causes a change of drift velocity of 12%. Tracks inclined of 45/degree/ have a resolution worse than those of 0/degree/ by a factor 3 /+-/ 2. A change in gas composition from CO/sub 2/(10%) to CO/sub 2/(11%) decreases the gas gain by 17 /+-/ 5%, and decreases drift velocity by 0.2 /+-/ 0.2%. The effect of an oxygen contamination of 3200 ppM is to change the mean pulse height by 45% over the 5 cm width of the cell. 4 refs., 15 figs.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.