Abstract

A technology-focused review of respiratory gas analysis, with an emphasis on carbon dioxide analysis, is presented. The measurement technologies deployed commercially are highlighted, and the basic principles and technical concerns of infrared spectroscopy and mainstream versus sidestream gas sampling are discussed. The specifications of particular interest to the clinician, accuracy and response time, and the related standard, with typical values for a capnometer, are presented. Representative time and volumetric capnograms are shown with the clinically relevant parameters described. Aspects of the terminology in present-day use and the need for clarity in defining what is a breath and an end-tidal value are reviewed. The applications of capnography of particular interest to the anesthesiologist are noted, and key references are provided. Ongoing developments with respect to respiratory gas analysis, and those that will impact it, are noted.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.