Abstract

The sensitivity of the resolving power of electron microscopes to ambient AC magnetic stray fields is well known. The highest permissible values of stray fields are fairly low and it may be difficult to find a suitable site for a microscope in normal laboratory conditions. A brief description of a device is given which reduces the disturbing effect of these stray fields by the method of automatic compensation. The device was tested on a JEM 6A transmission microscope. Even in an AC field as high as 24 mG (peak to peak) substantially higher resolving power was attained than that guaranteed by the manufacturer. Possible applications of the device to other electron beam instruments and in different magnetic stray fields are briefly discussed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.