Abstract

Continuous miniature crystal element (cMiCE) detectors are a potentially lower cost alternative for high resolution discrete crystal PET detector designs. We report on performance characteristics of a prototype PET scanner consisting of two cMiCE detector modules. Each cMiCE detector is comprised of a 50 × 50 × 8 mm3 LYSO crystal coupled to a 64 channel multi-anode PMT. The cMiCE detectors use a statistics-based positioning method based upon maximum likelihood estimation for event positioning. By this method, cMiCE detectors can also provide some depth of interaction event positioning information. For the prototype scanner, the cMiCE detectors were positioned across from one another on a horizontal gantry with a detector spacing of 10.7 cm. Full tomographic data were collected and reconstructed using single slice rebinning and filtered back projection with no smoothing. The average image resolutions in X (radial), Y (transverse) and Z (axial) were 1.05 ± 0.08 mm, 0.99 ± 0.07 mm, 1.24 ± 0.31 mm FWHM. These initial imaging results from a prototype imaging system demonstrate the outstanding image resolution performance that can be achieved using the potentially lower cost cMiCE detectors.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.