Abstract

The line spread function (LSF) of an imaging system incorporating a pinhole of 0.275 cm radius and a scintillation camera has been measured and its full width at half maximum (FWHM) compared to the values predicted by an expression based on the resolution index of Anger. In most cases, the actual resolution of the system is approximately 20% superior to that predicted by the above expression. However, good agreement is observed between our measured LSFs and those computed using a frequency domain method. Based on these computed LSFs, an empirical expression relating the FWHM of the total system to pinhole and camera parameters with an accuracy of +/- 3% is presented. This expression is applicable to most pinhole imaging systems in use today.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.