Abstract
Two-dimensional `acceptance diagrams' are used to discuss the transmission function and the resolution of neutron three-axis spectrometers. The discussion is simplified by assuming rectangular rather than the usual Gaussian transmission profiles for the beam elements,i.e.collimators and crystal mosaic spreads, but the results can be applied to the Gaussian profile approximation. The formalism clearly shows the effects of beam elements on the instrument transmission to the detector. Suitable choices of beam elements are determined to match the transmission functions of the primary and secondary spectrometers and hence optimize measurements. The effect of sample mosaic is addressed. The `focusing angle' for inelastic scattering is determined and discussed.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.