Abstract

We demonstrate dynamic spectral shaping in an optical coherence tomography (OCT) system with multiple scans of segmented spectral bands and software signal processing. With an effective spectral full-width at half-maximum (FWHM) of 45 nm, centered at 950 nm, we compress the interference fringe FWHM from 31 microns down to 7 microns, which is smaller than the theoretical limit of 8.9 microns if a Gaussian spectral shape is assumed, through the combined effect of dispersion compensation and spectral shaping.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.