Abstract

To compare: (1) the force required to initiate subsidence, and (2) the relative subsidence, of femoral stems implanted into intact femora, and then into the same femora in which an induced fissure had been stabilized by cerclage. In vitro, mechanical study. Femora (n=9) from 9 dogs. Femora were prepared for implantation of an uncemented stem. Stems were implanted with continuous and impact loading. After axial loading until a fissure occurred, the stems were extracted, and the fissure stabilized with double-loop cerclage. Stems were reimplanted, and reloaded to failure. Mean±SD load to initiate subsidence in intact femora was 1706±584 N compared with 2379±657 N for cerclaged bones (P=.002). Mean relative subsidence of intact femora was 3.99±2.09 mm compared with 1.79±2.99 mm for cerclaged bones (P=.091). The load to initiate subsidence is increased in femora that have fissured, then have been stabilized with double-loop cerclage, when compared with intact femora. The relative subsidence is not different between intact and stabilized specimens.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.