Abstract

Residual solvents in pharmaceutical samples are monitored using gas chromatography with head space. Based on good manufacturing practices, measuring residual solvents is mandatory for the release testing of all active pharmaceutical ingredients (API). The analysis of residual organic solvents (methanol, acetone, cyclohexane, dichloromethane, toluene) in Omeprazole, an active pharmaceutical ingredient was investigated. Omeprazole is a potent reversible inhibitor of the gastric proton pump H+/K+-ATPase. The Head space gas chromatography (HSGC) method described in this investigation utilized a SPB TM-624, Supelco, 30 m long x 0.25 mm internal diameter, 1.4µm-thick column. Since Omeprazole is a thermally labile compound, the selection of the proper injector temperature is critical to the success of the analysis. The injector temperature was set at 170ºC to prevent degradation. The initial oven temperature was set at 40ºC for 12 min and programmed at a rate of 10ºC min-1 to a final temperature of 220ºC for 5 min. Nitrogen was used as a carrier gas. The sample solvent selected was N,N-dimethylacetamide. The method was validated to be specific, linear, precise, sensitive, rugged and showed excellent recovery.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call