Abstract

For the detection and quantification of residual solvents in solutions of radiopharmaceuticals for human applications, a new analysis method was set up. With a simple direct gas chromatography injection, no special sample treatments or special injection techniques (such as head-space or solid-phase microextraction) are necessary. This method is especially suited for the fast and quantitative analysis of residual solvents in radiopharmaceutical formulations where 5-12% ethanol is present. The method proved to be both accurate and linear. The quantification limit is in the lower parts-per-million range, which is sufficient given the need for a fast analysis and the maximally allowed concentration for these residual solvents.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.