Abstract
For a magnetocardiography (MCG) system inside a magnetically shielded room (MSR), the residual field should be further suppressed to obtain high-quality MCG signals. In this paper, a compensation system has been developed based on the bi-planar coil and the proportional-integral-derivative (PID) controller. The bi-planar coil, derived from the target-field theory and the Tikhonov regularization method, is utilized to generate magnetic field with high uniformity in the pre-defined target region. And the PID controller is introduced to provide dynamic compensation current for the coil, according to the residual field change monitored by a reference SQUID magnetometer. The measurement results show that the noise suppression ratio (NSR) can reach above 20 dB in the low-frequency range from 0.1 Hz to 50 Hz. The DC component and the fluctuation of the residual field in time-domain can be respectively suppressed to 0 pT and 4 pT, indicating that this proposed compensation method is useful for the MCG measurement.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.