Abstract

In this study, we investigated the effects of various disinfectants used to prevent infectious diseases on medical images and medical equipment. First, we investigated the effect of residual disinfectant on medical images in CT, mammography (MMG), and general imaging systems. Acrylic discs with various disinfectants attached were photographed using each imaging device, and visual evaluation and changes in image signal values were evaluated. We also conducted a questionnaire survey of each manufacturer regarding cleaning methods for medical devices. With CT/MMG, residual disinfectant could be visually confirmed on the image. Although this could not be confirmed with the general imaging system, a significant difference was confirmed in the image signal values of the general imaging system through statistical analysis. This is thought to be largely due to the influence of nonlinearity in the short-time imaging range of general imaging equipment. In addition, from the responses to a questionnaire survey of each medical device manufacturer, we were able to understand detailed cleaning methods that are not covered in medical device instruction manuals.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.