Abstract

An efficient method was developed for the simultaneous determination of 10 industrial dyes (basic orange 2, basic orange 21, basic orange 22, acid orange II, auramine, basic rhodamine B and Sudan I-IV) in the foodstuffs using high-performance liquid chromatography coupled with diode array detector. Samples were extracted with acetonitrile and cleaned up on a solid phase extraction cartridge using HLB. The chromatographic separation was achieved on a C18 column using a mobile phase consisting of methanol and 10 mmol/L ammonium acetate with 0.1% formic acid by gradient elution. Good linearity (r ≥ 0.9993) was observed between 0.050 and 5.0 μg/mL. The limits of detection were in the range of 0.007-0.01 mg/kg, high recoveries (80.6-104%) and good reproducibility (1.1-5.7%) were obtained. Such method is simple, feasible and accurate, which can be applied to the quantification of 10 dyes in food samples.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.