Abstract
In our research, terahertz frequency-domain spectroscopy based on continuous-wave and coherent detection, which offers advantages of frequency selectivity and much higher frequency resolution, is used to acquire the optical parameter of nonpolar liquid samples in a polytetrafluoroethylene (PTFE) rectangle cell. The transfer matrix method is raised to remove the oscillation generated by the Fabry–Perot interferometer caused by the cell wall; thereby, dielectric constants of liquids can be extracted successfully. The refractive index of the nonpolar liquid samples cyclohexane ( C 6 H 12 ) and n -Hexane ( C 6 H 14 ) are obtained within the frequency range of 400–1600 GHz. The processed results of cyclohexane and n-Hexane in our method are basically stable with a refractive index of about 1.42 and 1.37, respectively, which agrees well with the experimental results using other methods previously.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.