Abstract

The non-source temperature calibration method of the multi-spectral pyrometer (MSP), put forward in the previous article, is the non-source calibration method based on the MSP transfer function (NCCTF). In this paper, the practicability of NCCTF is researched. The situations that maybe encountered of the temperature points in the actual calibration process are simulated. The impact of the intervals temperature points of different number and unequal intervals temperature points on the NCCTF accuracy are researched theoretically to determine the temperature points select method. Simulation experimental results show that the NCCTF can meet the actual calibration process with good usability.