Abstract
Objective To study the quality testing of dose delivery system of the active spot scanning proton and heavy ion accelerator, in order to provide the reference for the quality control of related equipment. Methods In the four therapy rooms, both 0.6 cc chambers and Gafchromic EBT3 films were used, respectively, to test the accelerator for dose reproducibility, dose linearity, dose stability, depth dose distribution, beam scanning position deviation and radiation field uniformity in each therapy room. Results Dose reproducibility variation coefficients are all less than 1.5%, dose linearity’s maximum deviations less than 2%, dose stability’s deviations less than 2%, depth dose distribution stability within 2%, beam scanning position deviation less than 1 mm, consistency of irradiation field’s deviation less than 2 mm, and flatness within ±5%. Conclusions The indicators about quality testing for the active spot scanning proton and heavy ion accelerator are all in line with the requirements of IEC standards draft. Key words: Proton and heavy ion accelerator; Active spot scanning; Quality test
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.