Abstract
To explore the high-risk fault risk of CT simulator and the main causes of the risk, and to put forward effective risk management strategies. The failure mode and effect analysis method was used to identify and control the operational fault risk of CT simulator. 5 major fault components, 8 fault failure models and 17 failure causes were analyzed. The top 5 failure causes are:anode target surface burn caused by direct scanning without warming up the tube (590.4), tube failure (518.2), burnout of joints caused by aging of high voltage cables (424.2), motor carbon brush wear (304.8) and belt break (296.4). The failure mode and effect analysis method can effectively identify the risk of equipment failure, and thus specifically formulate risk management and control measures to ensure the normal operation of equipment and the safety of doctors and patients.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Zhongguo yi liao qi xie za zhi = Chinese journal of medical instrumentation
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.