Abstract
The design and calibration of high current with high-stability has been a problem in metrology, a current calibration system with high-stability is introduced, which includes a novel DC current comparator, an original sampling method for high-precision current source, an adaptive control method for air enclosure. The stabilities of major components are analyzed and measured individually. The measurement results show that when the current is 300 A, the stability in one hour can be up to 5.7×10−7, and the uncertainty of measurement is 1.8×10−6.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.