Abstract
This article uses CAXA 3D solid design software to model the fracture reduction fixator, and assemble the model according to the requirements of clinical medicine. Obviously apply 50N, 100N, 150N, 200N axial load to the model, and obtain the relationship between axial load force and displacement. The stiffness of the model calculated by SPSS23.0 statistical analysis software is 88.5N/mm. The stiffness of the fracture reducer measured by the biomechanical model was 85.4N/mm, and the ratio between the two was 1.037, which was within the error range. Prove that the model and calculation method established in this study can provide reference for clinical medicine.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.