Abstract
In order to better simulate the flowing condition of wet gas, Tianjin University has designed and built up a wet gas flow experimental facility with adjustable and intermediate pressure in the flow laboratory. The designed pressure of the facility which used standard meter method and dual closed-loop design is 4MPa. The experiment medium is air and water, and the highest operation pressure is 1.6MPa. The gas flow rate range is 3~1000m3/h, and the liquid flow rate range is 0.05~8 m3/h. This article includes the structure introduction of the facility and the calculation of pressure loss of the system, etc. By the uncertainty assessment for the discussed facility, the conclusion is stated that the facility’s gas measurement uncertainty is 1% and the facility’s liquid measurement uncertainty is 0.35%.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.