Abstract

The problems of transfer Quantum Hall resistance to 1 Omega resistance standard with direct current comparator bridge (DCC) are introduced. Some effective technique ways are suggested by analyzing the possible causations of the measurement results' fluctuation. The comparing experiments between DCC and cryogenic current comparator bridge (CCC) have been performed with analyzing the uncertainty of measurement results. The transfer uncertainty from Quantum Hall resistance to 1 Omega standard with DCC is 3times10 <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">-8</sup> (k=1), which is enough accurate in practical resistance standard calibration of metrology.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.