Abstract

Owing to the geometric distortions of the X-ray images, C-arm CT (computed tomography) imaging system usually suffers from adverse effects. To overcome the difficulties, an integrated approach has been put forward to correct distorted X-ray images in the paper. The approach firstly uses OTSU algorithm and morphological open operation to extract the control points coordinates in the distorted X-ray images. To obtain more coordinates and improve the correction precision, least squares is used to fit the polynomial passing by the extracted points. More coordinates are calculated by using the fitted polynomial. Then, a biharmonic spline surface is interpolated by the extracted and calculated coordinates. Finally, the distorted X-ray images are corrected by using the interpolated biharmonic spline surface. Experiment on the acquired distorted X-ray images demonstrates that the proposed approach can effectively correct the distorted X-ray images. A comparison with classical approaches further shows that the approach proposed in the paper performs better in visual effects and correction accuracy.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.