Abstract
To develop an accurate, precise, specific high performance liquid chromatography method for quantification of Canagliflozin in bulk and dosage forms. A C18 column (250mm X 4.6mm; 5μm phenomenex) was used with mobile phase containing Acetonitrile-0.1% sodium acetate buffer (pH-4.6), (20:80) in isocratic mode. The flow rate maintained was 1.0ml/min and the U.V detector was operated at 291nm. The retention time of Canagliflozin was 3.307min and showed a good linearity in concentration range of 2-14μg/ml with correlation coefficient of 0.999. The average percent recovery was found to be 99.98%. The developed method follows validation parameters such as system suitability, linearity, precision, accuracy, limit of detection and limit of quantification and robustness as per ICH guidelinesQ2(R1). The proposed method was found to provide faster retention time with sharp resolution with linearity at a lowest concentration as compared to previous methods and this method is validated as per International conference on harmonization guidelines and successfully applied for bulk and pharmaceutical dosage form.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: SAMRIDDHI : A Journal of Physical Sciences, Engineering and Technology
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.