Abstract
As a kind of portable three-coordinate, the comparator is playing an more and more important role in precision industry measurement. Therefore, it is of great scientific significance and obvious socioeconomic benefits to carry out the research of comparator. This paper was based on rich experience with the institute in the field of the geometric measurement technology for a long time. And the measurement uncertainty of the instrument has been researched. The uncertainty of the diameter indication error is U=1.2μm k=2, the uncertainty of roundness indication error is U=1.4μm k=2, the uncertainty of sphere-center distance error is U=1.5μm+7.2×10−6 L k=2 L:m, and the uncertainty of length indication error is U=1.2μm+1.5×10−6 L k=2 L:m.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.