Abstract

In the control system of a quantum gravimeter, the optical signal containing phase information is weak and usually at the nV level. This situation puts forward high requirements for the measurement performance of the system. As the coupling input component of the measurement system, the noise level of the preamplifier has a dramatic influence on the noise performance. In this paper, the input current signal is provided by the photodiode. According to the photodiode output characteristics, we design three preamplifiers with different gain and bandwidth, which is verified by the optical system. Under the condition of meeting the design requirements, the noise performance is compared with the common commercial detector Model 2307. The results show that the noise performance of the photoelectric converter designed in this paper is about 1.5 times that of model 2307.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.