A new highly selective poly vinyl chloride (PVC) membrane thiocyanate (SCN–) electrode of copper (II) complex with ethylene-diamine-bis-(salicylaldehyde) [Cu (II)-SALEN] base with neutral carrier was studied, it displays a preferential potentiometric response to SCN– . The electrode exhibited near Nernst potential response to SCN– with a concentration range of 1.0×10-1~1.0×10-7 mol/L, a slope of -46 mV/p SCN– and a detection limit of 3.5×10-8 mol/L in pH5.5 of phosphate buffer solution at 25°C. PVC membrane electrode based on Cu (II)-SALEN had good selectivity to SCN– and some common anions had little interference with it. This electrode was applied to the determination of SCN– in waste water and the result obtained were in good agreement with those of high performance liquid chromatography (HPLC) method.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.