Abstract

35 patients with a smooth, threaded acetabular Mecron type prosthesis were examined with the aid of a table top with wiremarkers and a fixed 30-degree wedge to allow for reproducible positioning. Under fluoroscopic control, pelvic and spot films were made. The inter- and intraobserver variability of anteversion and inclination angle measurements of the prostheses had a standard deviation of less than 1 degree. The method can be applied to other acetabular prostheses as well.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.