Abstract

Increased breast density is a significant independent risk factor for breast cancer, and recent studies show that this risk is modifiable. Hence, breast density measures sensitive to small changes are desired. Utilizing fat-water decomposition MRI, we propose an automated, reproducible breast density measurement, which is nonionizing and directly comparable to mammographic density (MD). Retrospective study. The study included two sample sets of breast cancer patients enrolled in a clinical trial, for concordance analysis with MD (40 patients) and reproducibility analysis (10 patients). The majority of MRI scans (59 scans) were performed with a 1.5T GE Signa scanner using radial IDEAL-GRASE sequence, while the remaining (seven scans) were performed with a 3T Siemens Skyra using 3D Cartesian 6-echo GRE sequence with a similar fat-water separation technique. After automated breast segmentation, breast density was calculated using FraGW, a new measure developed to reliably reflect the amount of fibroglandular tissue and total water content in the entire breast. Based on its concordance with MD, FraGW was calibrated to MR-based breast density (MRD) to be comparable to MD. A previous breast density measurement, Fra80-the ratio of breast voxels with <80% fat fraction-was also calculated for comparison with FraGW. Pearson correlation was performed between MD (reference standard) and FraGW (and Fra80). Test-retest reproducibility of MRD was evaluated using the difference between test-retest measures (Δ1-2 ) and intraclass correlation coefficient (ICC). Both FraGW and Fra80 were strongly correlated with MD (Pearson ρ: 0.96 vs. 0.90, both P < 0.0001). MRD converted from FraGW showed higher test-retest reproducibility (Δ1-2 variation: 1.1% ± 1.2%; ICC: 0.99) compared to MD itself (literature intrareader ICC ≤0.96) and Fra80. The proposed MRD is directly comparable with MD and highly reproducible, which enables the early detection of small breast density changes and treatment response. 3 Technical Efficacy: Stage 1 J. Magn. Reson. Imaging 2018;48:971-981.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call