Abstract
In this study, the overall accuracy of normoxic polyacrylamide gel (nPAG) dosimeters is considered. Different badges of nPAG are fabricated, poured in containers of glass and Barex" and irradiated with a 6MV square photon beam. The polymer gel dosimeters were read out using MRI. The overall reproducibility and accuracy of nPAG gel dosimeters was determined by comparison with depth-dose profiles acquired with a Pin Point ionization chamber. Additionally, the effect of the container wall on the depth-dose profile and the effect of temperature changes before and after irradiation on the R2-dose response have been investigated. The average standard deviation and maximum deviation between 8 gel-measured depth-dose profiles and a depth-dose profile measured with an ionization chamber amounted to 0.543 Gy (2.5%) and 2.579 Gy (11.7%) respectively.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.