Abstract

BackgroundThe reproducibility of patient setup for radiotherapy is based on various methods including external markers, X-rays with planar or computerized image acquisition, and, more recently, surface matching imaging. We analyzed the setup reproducibility of 16 patients affected by prostate cancer who underwent conformal radiotherapy with curative intent by using a surface image registration system.MethodsWe analyzed the setup reproducibility of 16 patients affected by prostate cancer candidates for conformal radiotherapy by using a surface image registration system. At the initial setup, EPID images were compared with DRRs and a reference 3D surface image was obtained by the AlignRT system (Vision RT, London, UK). Surface images were acquired prior to every subsequent setup procedure. EPID acquisition was repeated when errors > 5 mm were reported.ResultsThe mean random and systematic errors were 1.2 ± 2.3 mm and 0.3 ± 3.0 mm along the X axis, 0.0 ± 1.4 mm and 0.5 ± 2.0 mm along the Y axis, and 2.0 ± 1.8 mm and -0.7 ± 2.4 mm along the Z axis respectively. The positioning error detected by AlignRT along the 3 axes X, Y, and Z exceeded the value of 5 mm in 14.1%, 2.0%, and 5.1% measurements and the value of 3 mm in 36.9%, 13.6% and 27.8% measurements, respectively. Correlation factors calculated by linear regression between the errors measured by AlignRT and EPID ranged from 0.77 to 0.92 with a mean of 0.85 and SD of 0.13. The setup measurements by surface imaging are highly reproducible and correlate with the setup errors detected by EPID.ConclusionSurface image registration system appears to be a simple, fast, non-invasive, and reproducible method to analyze the set-up alignment in 3DCRT of prostate cancer patients.

Highlights

  • The reproducibility of patient setup for radiotherapy is based on various methods including external markers, X-rays with planar or computerized image acquisition, and, more recently, surface matching imaging

  • Accurate and repeatable patient setup is a pre-requisite for radiotherapy in order to limit the margin around the clinical target volume (CTV), i.e. the planning target volume (PTV), and minimize the irradiation of healthy tissues responsible for early and late side effects

  • Both aspects have to be taken into account as prescribed by the ICRU 62 document (ICRU62) that defined in this regard the setup margin and the internal margin around the CTV to obtain the final PTV

Read more

Summary

Introduction

The reproducibility of patient setup for radiotherapy is based on various methods including external markers, X-rays with planar or computerized image acquisition, and, more recently, surface matching imaging. We analyzed the setup reproducibility of 16 patients affected by prostate cancer who underwent conformal radiotherapy with curative intent by using a surface image registration system. The reproducibility of external patient alignment is independent from the internal organ motion that can affect the position of the tumor with respect to the surrounding healthy tissues. Both aspects have to be taken into account as prescribed by the ICRU 62 document (ICRU62) that defined in this regard the setup margin and the internal margin around the CTV to obtain the final PTV. The system includes software designed to facilitate patient setup by surface-model acquisition and alignment by surface matching with a reference. In order to optimize the alignment process, the software is able to calculate the optimal rigid-body transformation (couch translation and rotation) that brings the surface model of the daily treatment fraction into congruence with the reference surface

Methods
Results
Discussion
Conclusion