Abstract
Background Coronary wall thickness measurements must be highly reproducible to be useful in longitudinal studies. While 2D turbo-spin-echo (TSE) and spiral techniques are commonly used, 3D techniques reduce partial-volume effects and allow greater vessel coverage. Recently, 3D coronary wall imaging was demonstrated with 100% respiratory efficiency (RE) using beat-to-beat respiratory motion correction (B2B-RMC) [1] which uses motion information determined from low resolution 3D images of the fat around the vessel as a surrogate for vessel motion. Here we assess the reproducibility of this technique together with that of navigator-gated 2D TSE and spiral acquisitions. Methods Cross-sectional right coronary artery wall images were obtained in 10 healthy subjects on a Siemens 1.5T Avanto scanner using dark-blood prepared B2B-RMC 3D spiral imaging, navigator-gated 2D TSE imaging and navigator-gated 2D spiral imaging. Acquisition order was randomized and subjects were imaged on two separate occasions to assess inter-study reproducibility using the intra-class correlation-coefficient (ICC) and Bland Altman analysis. All acquisitions used 0.7x0.7mm inplane resolution. B2B-RMC 3D spiral acquisitions acquired 8x3.0mm slices (16x1.5mm reconstructed) and 2D techniques acquired 1x6mm slice. Navigator-gated techniques used a 5mm window while B2B-RMC excluded data acquired at only very extreme respiratory positions, retrospectively correcting the rest. Durations, assuming 100% RE, were 600 cardiac cycles (CC) for B2B-RMC, 75CC for 2D spiral and 202-576CC for 2D
Highlights
Coronary wall thickness measurements must be highly reproducible to be useful in longitudinal studies
3D coronary wall imaging was demonstrated with 100% respiratory efficiency (RE) using beat-to-beat respiratory motion correction (B2B-RMC) [1] which uses motion information determined from low resolution 3D images of the fat around the vessel as a surrogate for vessel motion
Cross-sectional right coronary artery wall images were obtained in 10 healthy subjects on a Siemens 1.5T Avanto scanner using dark-blood prepared B2B-RMC 3D spiral imaging, navigator-gated 2D TSE imaging and navigator-gated 2D spiral imaging
Summary
Coronary wall thickness measurements must be highly reproducible to be useful in longitudinal studies. While 2D turbo-spin-echo (TSE) and spiral techniques are commonly used, 3D techniques reduce partial-volume effects and allow greater vessel coverage. 3D coronary wall imaging was demonstrated with 100% respiratory efficiency (RE) using beat-to-beat respiratory motion correction (B2B-RMC) [1] which uses motion information determined from low resolution 3D images of the fat around the vessel as a surrogate for vessel motion. We assess the reproducibility of this technique together with that of navigator-gated 2D TSE and spiral acquisitions
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.