Abstract
Blood pressure (BP) differences between two consecutive 24-h monitoring periods (P1, P2) were analyzed in 159 hospitalized pregnant women. Reproducibility index, or twice the standard deviation of differences between individual systolic and diastolic means, was better for 24-h (7.8 and 5.8) and daytime (8.7 and 6.3) than for nighttime (12.2 and 9.7) BP means. It did not depend on BP level or gestation week. Alert reaction to monitoring procedure increased BP only in the first 2 h of P1. The high overall reproducibility of ambulatory BP in pregnancy supports the expanding use of the technique also in this condition.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.