Abstract

BackgroundReprocessed (re-sterilized) bipolar energy devices represent one effort to reduce operative costs. MethodsBetween January 2014 to October 2015, 76 patients underwent laparoscopic colectomy using a reprocessed bipolar energy device and were case-matched to 76 patients from a prospectively-maintained database from November 2012 to December 2013 when an identical, new device was used. Outcomes included reprocessed device safety, efficiency and hospital costs. ResultsThere was no difference in patient demographics, operative times or failed pedicle ligation requiring intervention between groups (all P > 0.05). In 19.7% of reprocessed cases, the surgeon opened an additional new device after dissatisfaction with the reprocessed instrument. Operating room costs and total costs were less for the reprocessed device group (all P < 0.05). ConclusionReprocessed bipolar energy devices were associated with savings in operative expenses, however, larger studies are warranted due to the high surgeon dissatisfaction regarding safety concerns with the reprocessed equipment.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.